Technical Support

  • Operation Procedure-Veterinary Immunofluorescence Quantitative Analyzer 

    Test Mode 1 – Quick Test 1. Turn on the power switch, HV-FIA 3000 will load on the system and enter the initial interface. 2. Click“Test”menu on the main screen. After self-checking success, the system will enter the test interface. 3. Slide the Dip Switch to the left and make sure the indicator light is red. 4. Take ID Card out from the test kit, insert the Test ID Card, and then click“Read ID Card”. 

    66 2019-09-20
  • Veterinary Coagulation Analyzer 

    ​​​​​​​Test Mode 1 - General Items Test 1.Turn on the power switch, HV-COA 7000 will loan on the system and enter the initial interface. 2. Click "Test" menu on the main screen and enter the test interface. Take out the test cartridge from package.3. After inserting the cartridge into the cartridge holder,HV-COA 7000 will enter the heating mode "countdown 25 secs". 25 secs later, the sampling port shows green light indicating the sample adding operation. 

    26 2019-09-19
  • Operation Procedure-PARACUP

    1.Keep the cup upright and carefully unscrew the lid to avoid spilling of solution.2.Hold the lid of the cup and use a spoon to take about a spoonful of stool samples.3.Put the sample into the cup, stir well with the solution, and tighten the lid.4.Place test cup upside down on cup holder and let stand 15 minutes.5.Hold the cup upside down, unscrew the cap, and drop 1-2 drops of liquid onto the glass slide.6.Cover with coverslip and conduct microscope observation.

    19 2019-09-10